MTU high-precision turbidity meter

● Monitoring of water supply plant before, after, filtration, sedimentation and factory outlet monitoring ● Municipal water quality monitoring ● Industrial process discharge and recycling water quality monitoring, circulating cooling water monitoring, activated carbon filter water monitoring, membrane filtration water monitoring, boiler feedwater monitor

The MTU high-precision turbidity meter strictly follows the ISO7027 international standard and uses the 90 ° scattering principle for measurement. After focusing the light emitted from the sensor light source assembly, it enters the turbidimeter downward, and encounters the suspended particles in the sample to generate scattered light. The photodetector can detect the scattered light at an angle of 90 ° with the incident beam. The scattered light The value is proportional to the turbidity of the measured sample. The sensor is designed with a bubble removal device, which can remove the air bubbles entrained in the sample stream, thereby eliminating the most significant interference in low turbidity measurement.

● Built-in bubble elimination system, which can eliminate the interference of bubbles to the measured value to the greatest extent;

● Digital sensor supporting Modbus protocol, plug and play, strong anti-interference ability, cable length can reach hundreds of meters;

● Reliable LED light source with a lifetime of up to 100,000 hours;

● The cleaning and maintenance is very simple, and it is corrected once every three months;
● The meter can provide one isolated RS232 digital interface and one isolated RS485 digital interface, supporting standard Modbus ASCII and Modbus RTU protocols;
● Programmable relay output: Any relay can be set to high alarm, low alarm, and timer modes;
● With the function of automatic saving of historical data, it can store and query historical data information for 10 years.

Measurement performance range: 0 ~ 10NTU, 0 ~ 100NTU
Resolution: 0.001NTU, 0.01NTU
Accuracy: ± 5% or ± 0.015NTU of reading at 0 ~ 10NTU, whichever is greater;
± 5% or ± 0.15NTU of reading at 0 ~ 100NTU, whichever is greater, calibration: factory calibration, on-site calibration display: LCD liquid crystal display date, time, measurement value, historical trend line, etc. Chinese and English operation menu Optional data storage: automatic storage function, power-off data protection, can store and query 10 years of historical data on-site setting: completed by the transmitter button

Output analog output: One analog current output isolated 0/4 ~ 20mA, maximum load 750Ω, fault current output current optional relay output: 3 programmable relays, which can be set as alarm function and timing function Relay capacity: 2A, 250VAC
Digital interface: MODBUS RS232 / RS485, Profibus DP optional

AC power supply: 85 ~ 265VAC, 50Hz / 60Hz
DC power supply: 24VDC ± 10%

Physical performance transmitter size: 270 × 205 × 100mm
Sensor size: 379 × 310 × 243mm
Material: Transmitter: PC + ABS Sensor: ABS
Installation: flow-through installation

Environmental Performance Protection Level: Transmitter: IP65 Sensor: IP65
Working temperature: Transmitter: -20 ℃ ~ 60 ℃ Sensor: 0 ℃ ~ 40 ℃
Relative humidity: 0 ~ 95%, non-condensing